Share Email Print
cover

Proceedings Paper

Pulsed photoacoustic Doppler flowmetry using a cross correlation method
Author(s): J. Brunker; P. Beard
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The feasibility of making spatially resolved measurements of blood flow using pulsed photoacoustic Doppler techniques has been explored. Doppler time shifts were quantified via cross-correlation of pairs of photoacoustic waveforms generated within a blood-simulating phantom using pairs of laser light pulses. The photoacoustic waves were detected using a focussed or planar PZT ultrasound transducer. This approach was found to be effective for quantifying the linear motion of micron-scale absorbers imprinted on an acetate sheet moving with velocities in the range 0.15 to 1.50 ms-1. The effect of the acoustic spot diameter and the time separation between the laser pulses on measurement resolution and the maximum measurable velocity is discussed. The distinguishing advantage of pulsed rather than continuous-wave excitation is that spatially resolved velocity measurements can be made. This offers the prospect of mapping flow within the microcirculation and thus providing insights into the perfusion of tumours and other pathologies characterised by abnormalities in flow status.

Paper Details

Date Published: 23 February 2010
PDF: 8 pages
Proc. SPIE 7564, Photons Plus Ultrasound: Imaging and Sensing 2010, 756426 (23 February 2010); doi: 10.1117/12.841760
Show Author Affiliations
J. Brunker, Univ. College London (United Kingdom)
P. Beard, Univ. College London (United Kingdom)


Published in SPIE Proceedings Vol. 7564:
Photons Plus Ultrasound: Imaging and Sensing 2010
Alexander A. Oraevsky; Lihong V. Wang, Editor(s)

© SPIE. Terms of Use