A special CdZnTe (CZT) device on THM grown crystal has been developed. The device has different work function metals on opposite electrodes yet operates at room temperature like a conventional back-to-back symmetric MSM detector and not a one directional Schottky diode device. Aiming at creating a big breakthrough in CZT imaging device technology, the special CZT device presented in this study is capable of increasing the photopeak count by up to 50% compared to conventional CZT imaging device while maintaining good room temperature energy resolution by not significantly trading off detector leakage current. Pixel pad size and interpixel gap on a 20x20x5 mm3, 8x8 pixel pattern that result in optimum detector efficiency and interpixel resistance are presented. Sensitivity improvement impact on other device configuration will also be discussed. The design is highly practical, reliable and suitable for mass production.

Paper Details

Date Published: 17 September 2009
PDF: 17 pages
Proc. SPIE 7449, Hard X-Ray, Gamma-Ray, and Neutron Detector Physics XI, 744902 (17 September 2009); doi: 10.1117/12.828514
Show Author Affiliations
H. Chen, Redlen Technologies (Canada)
S. A. Awadalla, Redlen Technologies (Canada)
P. Marthandam, Redlen Technologies (Canada)
K. Iniewski, Redlen Technologies (Canada)
P. H. Lu, Redlen Technologies (Canada)
G. Bindley, Redlen Technologies (Canada)


Published in SPIE Proceedings Vol. 7449:
Hard X-Ray, Gamma-Ray, and Neutron Detector Physics XI
Ralph B. James; Larry A. Franks; Arnold Burger, Editor(s)

© SPIE. Terms of Use
Back to Top