Performance evaluation of a full line of medical diagnostic displays and test of a web-based service for remote calibration and quality assurance
Author(s): S. Busoni; G. Belli; A. Taddeucci; S. Targetti; I. Menchi; C. Gori; L. Capaccioli; A. Falivene; G. Spanò; A. Guasti; P. Saletti; F. Rossi
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

Main goal of this study is to compare the performances of a full line of LCD diagnostic display systems, in terms of white point luminance, accuracy and stability over time, GSDF conformance and luminance uniformity, and to test a web-based service for remote calibration and QA in a large hospital. The display systems under test included 3MP and 5MP grayscale, 2MP and 6MP color and 5MP mammography LCD monitors, all manufactured by BARCO, for a total amount of 119 units. Measured performances were all within the acceptance range proposed by the major international protocols and show a very good stability in time, except for a few cases. The web-based service for remote QA and calibration resulted well suited for the management of a large scale medical facility, where high performance displays are in use and time saving QA programs and a central QA policy are both needed.

Paper Details

Date Published: 12 March 2009
PDF: 12 pages
Proc. SPIE 7263, Medical Imaging 2009: Image Perception, Observer Performance, and Technology Assessment, 72630A (12 March 2009); doi: 10.1117/12.812283
Show Author Affiliations
S. Busoni, Azienda Ospedaliero-Universitaria Careggi (Italy)
G. Belli, Azienda Ospedaliero-Universitaria Careggi (Italy)
A. Taddeucci, Azienda Ospedaliero-Universitaria Careggi (Italy)
S. Targetti, Azienda Ospedaliero-Universitaria Careggi (Italy)
I. Menchi, Azienda Ospedaliero-Universitaria Careggi (Italy)
C. Gori, Azienda Ospedaliero-Universitaria Careggi (Italy)
L. Capaccioli, Azienda Ospedaliero-Universitaria Careggi (Italy)
A. Falivene, Azienda Ospedaliero-Universitaria Careggi (Italy)
G. Spanò, Azienda Ospedaliero-Universitaria Careggi (Italy)
A. Guasti, Azienda Ospedaliero-Universitaria Careggi (Italy)
P. Saletti, Azienda Ospedaliero-Universitaria Careggi (Italy)
F. Rossi, Azienda Ospedaliero-Universitaria Careggi (Italy)


Published in SPIE Proceedings Vol. 7263:
Medical Imaging 2009: Image Perception, Observer Performance, and Technology Assessment
Berkman Sahiner; David J. Manning, Editor(s)

© SPIE. Terms of Use