Proceedings Paper
High-speed spectral-domain optical coherence tomography at 830nm wavelengthFormat | Member Price | Non-Member Price |
---|---|---|
$14.40 | $18.00 |
![]() |
GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. | Check Access |
Paper Abstract
We present a high-speed spectral domain optical coherence tomography (SD-OCT) system at 830
nm wavelength which is consisted of a fiber based Michelson interferometer and a custom-built
spectrometer. The designed resolution of the spectrometer is about 67.4pm which limits the
maximum detection depth 2.56mm in air. And the 35us exposure time of the high speed line scan
CCD makes real-time imaging possible. Furthermore, a novel method of spectrometer calibration
is put forward. The method can remove the influence of dispersion mismatch, thus accurately
determine the distribution of wavelength on the line scan CCD, which leads to a precise
interpolation and a subsequent better contrast image.
Paper Details
Date Published: 8 January 2008
PDF: 9 pages
Proc. SPIE 6826, Optics in Health Care and Biomedical Optics III, 68260A (8 January 2008); doi: 10.1117/12.757208
Published in SPIE Proceedings Vol. 6826:
Optics in Health Care and Biomedical Optics III
Xingde Li; Qingming Luo; Ying Gu, Editor(s)
PDF: 9 pages
Proc. SPIE 6826, Optics in Health Care and Biomedical Optics III, 68260A (8 January 2008); doi: 10.1117/12.757208
Show Author Affiliations
Published in SPIE Proceedings Vol. 6826:
Optics in Health Care and Biomedical Optics III
Xingde Li; Qingming Luo; Ying Gu, Editor(s)
© SPIE. Terms of Use