We have investigated the application of narrow-band intracavity spectroscopy for high throughput characterization and detection of reaction products in-situ. An external cavity diode laser (ECDL), tunable from 810-890 nm, is used to measure varying concentrations of a range of compounds in solution. A linear calibration curve for parts per billion (ppb) concentrations corresponding to 17 nM has been achieved, levels not detectable with UV/VIS spectrometers. By replacing the grating with a high reflective mirror, parts per trillion (ppt) concentration detection was also achieved corresponding to 340 pM. We believe that this technique provides a method for fast and safe measurements of reactions in real time and has potential applications in the pharmaceutical and chemical industry as well as for integrated bio-diagnostics lab-on-a-chip devices.

Paper Details

Date Published: 10 February 2007
PDF: 9 pages
Proc. SPIE 6444, Ultrasensitive and Single-Molecule Detection Technologies II, 644407 (10 February 2007); doi: 10.1117/12.696558
Show Author Affiliations
Untzizu Elejalde, Univ. of Strathclyde (United Kingdom)
John M. Girkin, Univ. of Strathclyde (United Kingdom)


Published in SPIE Proceedings Vol. 6444:
Ultrasensitive and Single-Molecule Detection Technologies II
Jorg Enderlein; Zygmunt K. Gryczynski, Editor(s)