Share Email Print
cover

Proceedings Paper

UV radiation dose measurements
Author(s): R. Cardoso; J. P. Valdez-Chaparro; E. Rosas
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

We present the implementation of the ultraviolet (UV) radiation dose measurement system at the Laboratorio de Dosimetria Optica (LDOS) of the Centro Nacional de Metrologia (CENAM). The UV dose measurement ranges established so far cover from 12 Jm-2 to 7 256 Jm-2 and from 171 Jm-2 to 102 842 Jm-2, for 254 nm and 365 nm, respectively; corresponding to two emission lines of an arc Hg(Xe) lamp. Those dose ranges can be reached with exposure times from 2 s to 10 min and the reference dose uncertainty has been estimated to be of ±5,4 %.

Paper Details

Date Published: 10 February 2006
PDF: 6 pages
Proc. SPIE 6046, Fifth Symposium Optics in Industry, 60461R (10 February 2006); doi: 10.1117/12.674596
Show Author Affiliations
R. Cardoso, Ctr. Nacional de Metrología (Mexico)
J. P. Valdez-Chaparro, Ctr. Nacional de Metrología (Mexico)
E. Rosas, Ctr. Nacional de Metrología (Mexico)


Published in SPIE Proceedings Vol. 6046:
Fifth Symposium Optics in Industry
Eric Rosas; Rocío Cardoso; Juan C. Bermudez; Oracio Barbosa-García, Editor(s)

© SPIE. Terms of Use
Back to Top