Share Email Print
cover

Proceedings Paper

Evaluating dynamic contrast-enhanced and photoacoustic CT to assess intra-tumor heterogeneity in xenograft mouse models
Author(s): Keith M. Stantz; Bo Liu; Minsong Cao; Dan Reinecke; Mario Dzemidzic; Yun Liang; Robert Kruger
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Purpose: To evaluate photoacoustic CT spectroscopy (PCT-S) and dynamic contrast-enhanced CT (DCE-CT) ability to measure parameters - oxygen saturation and vascular physiology - associated with the intra-tumor oxygenation status. Material and Methods: Breast (VEGF165 enhance MCF-7) and ovarian (SKOV3x) cancer cells were implanted into the fat pads and flanks of immune deficient mice and allowed to grow to a diameter of 8-15 mm. CT was used to determine physiological parameters by acquiring a sequence of scans over a 10 minute period after an i.v. injection of a radio-opaque contrast agent (Isovue). These time-dependent contrast-enhanced curves were fit to a two-compartmental model determining tumor perfusion, fractional plasma volume, permeability-surface area produce, and fractional interstitial volume on a voxel-by-voxel basis. After which, the tumors were imaged using photoacoustic CT (Optosonics, Inc., Indianapolis, IN 46202). The near infrared spectra (700-910 nm) within the vasculature was fit to linear combination of measured oxy- and deoxy-hemoglobin blood samples to obtain oxygen saturation levels (SaO2). Results: The PCT-S scanner was first calibrated using different samples of oxygenated blood, from which a statistical error ranging from 2.5-6.5% was measured and a plot of the hemoglobin dissociation curve was consistent with empirical formula. In vivo determination of tumor vasculature SaO2 levels were measurably tracked, and spatially correlated to the periphery of the tumor. Tumor depend variations in SaO2 - 0.32 (ovarian) and 0.60 (breast) - and in vascular physiology - perfusion, 1.03 and 0.063 mL/min/mL, and fractional plasma volume, 0.20 and 0.07 - were observed. Conclusion: Combined, PCT-S and CED-CT has the potential to measure intra-tumor levels of tumor oxygen saturation and vascular physiology, key parameters associated with hypoxia.

Paper Details

Date Published: 20 March 2006
PDF: 12 pages
Proc. SPIE 6143, Medical Imaging 2006: Physiology, Function, and Structure from Medical Images, 61431F (20 March 2006); doi: 10.1117/12.654056
Show Author Affiliations
Keith M. Stantz, Purdue Univ. (United States)
Indiana Univ. (United States)
Bo Liu, Purdue Univ. (United States)
Minsong Cao, Purdue Univ. (United States)
Dan Reinecke, Optosonics, Inc. (United States)
Mario Dzemidzic, Indiana Univ. (United States)
Yun Liang, Indiana Univ. (United States)
Robert Kruger, Optosonics, Inc. (United States)


Published in SPIE Proceedings Vol. 6143:
Medical Imaging 2006: Physiology, Function, and Structure from Medical Images
Armando Manduca; Amir A. Amini, Editor(s)

© SPIE. Terms of Use