Novel SNR determination method in parallel MRI
Author(s): Tosiaki Miyati; Hiroshi Imai; Akio Ogura; Tsukasa Doi; Toshio Tsuchihashi; Yoshio Machida
Format Member Price Non-Member Price
PDF $17.00 $21.00

Paper Abstract

We devised two kinds of new methods for accurate measurement of the image signal-to-noise ratio (SNR) in parallel magnetic resonance imaging (MRI) because image noise of the parallel MRI was not spatially constant. Using the first (Consecutive) method, more than fifty consecutive scans of the uniform phantom were obtained with identical scan parameters. Then the SNRs in each pixel were calculated from the ratio of mean signal intensity to the standard deviation of the time domain on a pixel-by-pixel. With the second (Remove) method, the phantom was removed after the first scan, and the second scan was done with identical parameters and the RF coil loading device. The SNRs in each pixel were then obtained from the ratio of the signal intensity of the first scan to the second scan (w/o phantom) image which was multiplied by the square root of 2/pi and filtered by the running mean (7 by 7 pixels). Moreover, actual geometry factors were calculated from image SNRs of parallel and no parallel MRI. The image SNR and actual geometry factor of parallel MRI with the Consecutive method agreed with that of the Remove method. The SNRs of the no parallel MRI with the above two methods conformed with that of the conventional SNR method (NEMA standard). Both new methods make it possible to obtain a more detailed determination of SNR in parallel MRI, and to calculate the actual geometry factor.

Paper Details

Date Published: 2 March 2006
PDF: 7 pages
Proc. SPIE 6142, Medical Imaging 2006: Physics of Medical Imaging, 61423O (2 March 2006); doi: 10.1117/12.653482
Show Author Affiliations
Tosiaki Miyati, Graduate School of Medicine, Kanazawa Univ. (Japan)
Hiroshi Imai, Anjo Kosei Hospital (Japan)
Akio Ogura, Kyoto City Hospital (Japan)
Tsukasa Doi, Nara Medical Univ. Hospital (Japan)
Toshio Tsuchihashi, Nippon Medical School Hospital (Japan)
Yoshio Machida, Toshiba Medical Systems Corp. (Japan)


Published in SPIE Proceedings Vol. 6142:
Medical Imaging 2006: Physics of Medical Imaging
Michael J. Flynn; Jiang Hsieh, Editor(s)

© SPIE. Terms of Use
Back to Top