Share Email Print
cover

Proceedings Paper

Frequency characteristics of images measured on the VolumeScope scanning device
Author(s): V. N. Vasiliev; M. Yu. Zaytsev
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The modulation transfer function was determined on the VolumeScope, a prototype X-ray Compton scatter tomographic scanner. The linear spread function of the device was measured independently along each coordinate axis with a 0.3 mm aluminum foil slightly angled with respect to the voxel grid. A technique of the presampled MTF generation was used to increase a sampling frequency of the data and reduce a statistical noise. The LSFs and MTFs along three coordinate axes are presented. The FWHMs of the LSF peaks are 2.50, 2.3 1 and 3.72 mm along the axes X, Y and Z respectively. The MTF cutoff frequencies are 0.45,0.50 and 0.35 cycle/mmrespectively.

Paper Details

Date Published: 6 December 2006
PDF: 4 pages
Proc. SPIE 5943, X-ray and Neutron Capillary Optics II, 59430V (6 December 2006); doi: 10.1117/12.637992
Show Author Affiliations
V. N. Vasiliev, Russian Scientific Ctr. of Roentgenology and Radiology (Russia)
M. Yu. Zaytsev, Institute for Roentgen Optics (Russia)


Published in SPIE Proceedings Vol. 5943:
X-ray and Neutron Capillary Optics II
Muradin A. Kumakhov; Richard B. Hoover, Editor(s)

© SPIE. Terms of Use
Back to Top