We present an imaging system which could be used for endoscopic topography. Indeed it allows for two dimensional information transmission through a unidimensional imaging channel which is a monomode optical fiber. The principle is the coupling of wavelength multiplexing and spectral interferometry and a special configuration renders this system dispersion self compensated what enables a high signal stability. Principles will be presented as well as results and limits of the system