Share Email Print
cover

Proceedings Paper

Coded excitation ultrasound system based on modified hermite series
Author(s): Fahmi Cheikhrouhou; Heinrich Martin Overhoff; Tobias Gehrke
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

Conventional ultrasound systems acquire an image by scanning the environment successively line by line. This limits the frame rate. Furthermore the used conventional signal affects the resolution and the penetration depth. In this contribution we propose to use modified mutually orthonormal Hermite series in a coded excitation system yielding a full image in only one transmission. Therefore a higher frame rate can be achieved. Additional properties of these codes offer the possibility to increase the resolution and the penetration depth. Preliminary results show the effectiveness of the proposed approach.

Paper Details

Date Published: 12 April 2005
PDF: 10 pages
Proc. SPIE 5750, Medical Imaging 2005: Ultrasonic Imaging and Signal Processing, (12 April 2005); doi: 10.1117/12.596213
Show Author Affiliations
Fahmi Cheikhrouhou, Univ. of Applied Sciences Gelsenkirchen (Germany)
Heinrich Martin Overhoff, Univ. of Applied Sciences Gelsenkirchen (Germany)
Tobias Gehrke, Univ. of Applied Sciences Gelsenkirchen (Germany)


Published in SPIE Proceedings Vol. 5750:
Medical Imaging 2005: Ultrasonic Imaging and Signal Processing
William F. Walker; Stanislav Y. Emelianov, Editor(s)

© SPIE. Terms of Use