Proceedings Paper

Investigation of cone-beam acquisitions implemented using a novel dedicated mammotomography system with unique arbitrary orbit capability
Author(s): Randolph L. McKinley; Caryl N. Bryzmialkiewicz; Priti Madhav; Martin P. Tornai
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

We investigate cone-beam acquisitions implemented on a novel dedicated cone-beam transmission computed mammotomography (CmT) system with unique arbitrary orbit capability for pendant, uncompressed breasts. We use a previously reported optimized quasi-monochromatic beam technique together with orbits made possible with a novel CmT gantry system, to evaluate Vertical-Axis-Of-Rotation (VAOR), Circle-Plus-Two-Arcs (CP2A), and Saddle trajectories. Aquisition parameters include: W target, 60 kVp tube potential, 100th VL Nd filtration, 1.25 mAs, 55 cm SID, CsI(Tl) digital flat panel x-ray detector, and 7.7cm diameter uniform disc (Defrise) and resolution phantoms. Complex orbits were also performed for a realistic breast phantom. Reconstructions used an iterative ordered subsets transmission (OSTR) algorithm with 4x4 binned projections, 8 subsets, and 10 iterations, with 0.125 mm3 voxels. We evaluate the results for image artifacts, distortion, and resolution. Reconstructed images of the disc coronal and sagittal slices show significant distortion of the discs and phantom interfaces away from the central plane of the cone-beam for VAOR, less distortion for CP2A, and minimal distortion for the complex 3D Saddle orbit. Resolution phantoms indicate no loss of resolution with the Saddle orbit, with the smallest 1.1mm diameter rods clearly resolved. Other image artifacts such as streaking were also significantly reduced in the Saddle orbit case. Results indicate that arbitrary orbits of pendant uncompressed breasts using cone-beam acquisitions and OSTR iterative reconstructions can be successfully implemented for dedicated CmT to improve angular sampling with significant reduction in distortion and other image artifacts. This capability has the potential to improve the performance of dedicated CmT by adequately sampling the breast and anterior chest volumes of prone patients with pendant, uncompressed breasts.

Paper Details

Date Published: 20 April 2005
PDF: 9 pages
Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); doi: 10.1117/12.596089
Show Author Affiliations
Randolph L. McKinley, Duke Univ. Medical Ctr. (United States)
Duke Univ. (United States)
Caryl N. Bryzmialkiewicz, Duke Univ. Medical Ctr. (United States)
Duke Univ. (United States)
Priti Madhav, Duke Univ. Medical Ctr. (United States)
Duke Univ. (United States)
Martin P. Tornai, Duke Univ. Medical Ctr. (United States)
Duke Univ. (United States)


Published in SPIE Proceedings Vol. 5745:
Medical Imaging 2005: Physics of Medical Imaging
Michael J. Flynn, Editor(s)

© SPIE. Terms of Use
Back to Top