Recently, a 3D filtered-backprojection (FBP)-based algorithm for image reconstruction on PI-line segments in a helical cone-beam CT scan has been developed (Zou and Pan, 2004). In the present work, we derive new reconstruction algorithms for circular cone-beam scans based upon this algorithm and a concept of virtual PI-line. We prove that, in the case of conventional full- and short-scan, the newly derived algorithms are mathematically identical to existing algorithms. More importantly, in the case of reduced-scans in which the scanning angle range is less than that in a short-scan, the new algorithms can yield exact region of interest (ROI) reconstruction in mid-plane and approximate ROI reconstruction in off-mid-planes. We have performed a preliminary numerical study that verifies our theoretical assertions.

Paper Details

Date Published: 29 April 2005
PDF: 6 pages
Proc. SPIE 5747, Medical Imaging 2005: Image Processing, (29 April 2005); doi: 10.1117/12.595857
Show Author Affiliations
Lifeng Yu, Univ. of Chicago (United States)
Dan Xia, Univ. of Chicago (United States)
Yu Zou, Univ. of Chicago (United States)
Xiaochuan Pan, Univ. of Chicago (United States)