Evaluation of an optically coupled CCD digital radiography system
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

A digital radiography system comprised of a large field of view (43x43cm) high luminance CsI scintillator, optically coupled to a 4096x4096 element CCD sensor with 12:1 demagnification was evaluated by measuring the modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE). Detector evaluation was performed using IEC standard #62220-1 methodologies for beam quality. In this study, RQA-5 (21 mm Al added filtration, 74 kVp, 7.1 mm half value layer (HVL)) and RQA-9 (40 mm Al added filtration, 119 kVp, 11.5 mm HVL) qualities were used at several incident exposures from <0.1 mR to >50 mR. Two detector modes of operation included high resolution (HR) and high efficiency (HE), with 108 and 216 μm pixel dimensions, respectively. The detector system responded to 60 mR incident exposure prior to saturation for the HR mode and up to 30 mR in the HE mode. The pre-sampled MTF(f) had 50% modulation at 0.95 mm-1 (HR) and 0.85 mm-1 (HE); and 10%MTF(f) was reached at 2.4 mm-1 (HR) and 2.0 mm-1 (HE). At a frequency of 0.5 mm-1, the DQE was 40% to 50%, and at 1 mm-1 was 12% to 20% for HR and HE modes, respectively. The DQE at low exposures was substantially better for the HE mode. Little dependence of the DQE on beam energy was found, but the RQA 9 beam had lower values. Above a frequency of 2 mm-1 the DQE dropped to zero, attributed to low MTF. Results suggest that patient radiation exposures equivalent or better than a conventional 400 speed screen-film detector can be achieved for many imaging procedures with sufficient SNR and spatial resolution required for a wide range of diagnostic radiography applications.

Paper Details

Date Published: 20 April 2005
PDF: 10 pages
Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); doi: 10.1117/12.595751
Show Author Affiliations
J. Anthony Seibert, Univ. of California/Davis (United States)
Alexander Kwan, Univ. of California/Davis (United States)
John M. Boone, Univ. of California/Davis (United States)
Paul Brown, Oregon Health & Science Univ. (United States)
Robin Winsor, Imaging Dynamics Co. (Canada)


Published in SPIE Proceedings Vol. 5745:
Medical Imaging 2005: Physics of Medical Imaging
Michael J. Flynn, Editor(s)

© SPIE. Terms of Use
Back to Top