Share Email Print
cover

Proceedings Paper

Comparison of air space measurement imaged by CT, small-animal CT, and hyperpolarized Xe MRI
Author(s): Aniseh Madani; Steven White; Giles Santyr; Ian Cunningham
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Lung disease is the third leading cause of death in the western world. Lung air volume measurements are thought to be early indicators of lung disease and markers in pharmaceutical research. The purpose of this work is to develop a lung phantom for assessing and comparing the quantitative accuracy of hyperpolarized xenon 129 magnetic resonance imaging (HP 129Xe MRI), conventional computed tomography (HRCT), and highresolution small-animal CT (μCT) in measuring lung gas volumes. We developed a lung phantom consisting of solid cellulose acetate spheres (1, 2, 3, 4 and 5 mm diameter) uniformly packed in circulated air or HP 129Xe gas. Air volume is estimated based on simple thresholding algorithm. Truth is calculated from the sphere diameters and validated using μCT. While this phantom is not anthropomorphic, it enables us to directly measure air space volume and compare these imaging methods as a function of sphere diameter for the first time. HP 129Xe MRI requires partial volume analysis to distinguish regions with and without 129Xe gas and results are within %5 of truth but settling of the heavy 129Xe gas complicates this analysis. Conventional CT demonstrated partial-volume artifacts for the 1mm spheres. μCT gives the most accurate air-volume results. Conventional CT and HP 129Xe MRI give similar results although non-uniform densities of 129Xe require more sophisticated algorithms than simple thresholding. The threshold required to give the true air volume in both HRCT and μCT, varies with sphere diameters calling into question the validity of thresholding method.

Paper Details

Date Published: 20 April 2005
PDF: 6 pages
Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); doi: 10.1117/12.594915
Show Author Affiliations
Aniseh Madani, Robarts Research Institute (Canada)
The Univ. of Western Ontario (Canada)
Steven White, Carleton Univ. (Canada)
Giles Santyr, Robarts Research Institute (Canada)
Carleton Univ. (Canada)
Ian Cunningham, Robarts Research Institute (Canada)
The Univ. of Western Ontario (Canada)


Published in SPIE Proceedings Vol. 5745:
Medical Imaging 2005: Physics of Medical Imaging
Michael J. Flynn, Editor(s)

© SPIE. Terms of Use