New fluorescence-based disposable probes have been developed for monitoring of pH, carbon dioxide, and oxygen in extracorporeal loops and radial arteries. These low cost, small diameter sensors fit through most standard female Luer Lock fittings. The portable monitor and its interface to the probes are described. In-vitro performance of the sensor systems versus tonometered bovine blood and blood analyzer calibration ampules are shown, along with a brief discussion of clinical applications of these devices.

Paper Details

Date Published: 16 April 1992
PDF: 10 pages
Proc. SPIE 1648, Fiber Optic Medical and Fluorescent Sensors and Applications, (16 April 1992); doi: 10.1117/12.58306
Show Author Affiliations
Lawrence S. Ring, Lightsense Corp. (United States)
Wayne E. Rodgers, Lightsense Corp. (United States)


Published in SPIE Proceedings Vol. 1648:
Fiber Optic Medical and Fluorescent Sensors and Applications
Douglas R. Hansmann; Fred P. Milanovich; Gerald G. Vurek; David R. Walt, Editor(s)