Share Email Print
cover

Proceedings Paper

Real-time photoacoustic imaging based on multielement linear transducer array
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

A real-time photoacoustic (PA) imaging system based on multi-element linear transducer array was developed and test on phantoms. A Q switched Nd:YAG laser operating at 532 nm and 20Hz repeat rate was used in our experiment as thermal source. The multi-element linear transducer array consists of 320 elements. By phase-controlled method, 64 signals, one of which gathered by 11-group element, make up of an image. It was acquired in only about 3 seconds. Phantom experiment results can map the distribution of the optical absorption correctly. Compared to other existing technology and algorithm, the PA imaging based on transducer array was characterize by speediness and convenience. It can provide a new approach for tissue functional imaging in vivo, and may have potentials in developing into an appliance for clinic diagnosis.

Paper Details

Date Published: 18 January 2005
PDF: 9 pages
Proc. SPIE 5630, Optics in Health Care and Biomedical Optics: Diagnostics and Treatment II, (18 January 2005); doi: 10.1117/12.573521
Show Author Affiliations
Bangzheng Yin, South China Normal Univ. (China)
Da Xing, South China Normal Univ. (China)
Yi Wang, South China Normal Univ. (China)
Yaguang Zeng, South China Normal Univ. (China)
Yi Tan, South China Normal Univ. (China)


Published in SPIE Proceedings Vol. 5630:
Optics in Health Care and Biomedical Optics: Diagnostics and Treatment II
Britton Chance; Mingzhe Chen; Arthur E. T. Chiou; Qingming Luo, Editor(s)

© SPIE. Terms of Use
Back to Top