Beam hardening correction and quantitative micro-CT
Author(s): Sven Prevrhal
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

The main use of micro-CT is to analyze bone structure, and values of the linear attenuation coefficients μ at every voxel position in the scanned volume are only used for the analysis of bone structure by establishing a threshold that separates bone from non-bone material. To additionally quantify the degree of mineralization of bone (DMB) from μ in multi-component samples, we corrected for beam hardening and its associated errors in DMB quantification caused by the polychromatic spectra of X-ray tubes used in bench-top micro-CT devices. The correction was implemented by simulating the difference of mono-chromatic and poly-chromatic X-ray sources and adding these differences to the original image in an iterative fashion. When applied to simple cylinders containing a single material, improvement on the constancy of the reconstructed voxel values could be observed to over 90% accuracy.

Paper Details

Date Published: 26 October 2004
PDF: 10 pages
Proc. SPIE 5535, Developments in X-Ray Tomography IV, (26 October 2004); doi: 10.1117/12.559976
Show Author Affiliations
Sven Prevrhal, Univ. of California/San Francisco (United States)