Share Email Print
cover

Proceedings Paper

Development of a high frequency (35-MHz) linear ultrasonic array
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

Methods for fabricating and modeling of a high frequency linear ultrasonic array are presented. This array was designed primarily for human eye imaging, and features elements mechanically diced out of a fine grain high density PZT-5H ceramic. Array elements were spaced with a 50 micron pitch, interconnected via a flexible circuit and matched to the 50 Ohm system electronics via a 85 Ohm transmission line coaxial cable. The current array design was based upon tradeoffs between the time domain response bandwidth sensitivity, as well as, the level of array encapsulation and suppression acoustic crosstalk. A prototype four element array was constructed with promising results. An average center frequency of 34 MHz with a -6 dB bandwidth of at least 45% per element was achieved with a 20 dB pulse length of 105 ns. The array performance was compared to a time domain finite element analysis program and excellent agreement between theory and experiment was achieved.

Paper Details

Date Published: 28 April 2004
PDF: 6 pages
Proc. SPIE 5373, Medical Imaging 2004: Ultrasonic Imaging and Signal Processing, (28 April 2004); doi: 10.1117/12.537062
Show Author Affiliations
Jonathan M. Cannata, Univ. of Southern California (United States)
K. Kirk Shung, Univ. of Southern California (United States)


Published in SPIE Proceedings Vol. 5373:
Medical Imaging 2004: Ultrasonic Imaging and Signal Processing
William F. Walker; Stanislav Y. Emelianov, Editor(s)

© SPIE. Terms of Use