Share Email Print
cover

Proceedings Paper

Ultrahigh-sensitivity wideband Fabry-Perot ultrasound sensors as an alternative to piezoelectric PVDF transducers for biomedical photoacoustic detection
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

A high sensitivity, wideband ultrasound sensor based on a high finesse Fabry-Perot (FP) polymer film interferometer has been demonstrated with a bandwidth of 20 MHz and a 50μm diameter active area. Used in conjunction with a balanced photodetector to enable the use of a high intensity interrogating light beam of up to 36mW, the sensor system provided a noise equivalent pressure (NEP) of 0.35kPa over a 20 MHz measured bandwidth. It is shown further that this NEP could, in principle, be reduced to 0.16 kPa by using an interrogating source with a wider wavelength tuning range than was available in the current study to track drift in the phase bias of the FP sensor. The sensitivity achieved is an order of magnitude higher than previously demonstrated with this type of sensors, and is comparable to that of a 1mm diameter PVDF element. The combination of high sensitivity and the small active area (<50μm diameter) makes the FP sensor scheme particularly suitable for photoacoustic imaging applications.

Paper Details

Date Published: 12 July 2004
PDF: 8 pages
Proc. SPIE 5320, Photons Plus Ultrasound: Imaging and Sensing, (12 July 2004); doi: 10.1117/12.531160
Show Author Affiliations
Edward Z. Y. Zhang, Univ. College London (United Kingdom)
Paul C. Beard, Univ. College London (United Kingdom)


Published in SPIE Proceedings Vol. 5320:
Photons Plus Ultrasound: Imaging and Sensing
Alexander A. Oraevsky; Lihong V. Wang, Editor(s)