Previously, we proposed a polarimetric method, that exploits the Brewster-reflection with the final goal of application to the human eye (reflection off the eye lens) for non-invasive glucose sensing. The linearly polarized reflected light of this optical scheme is rotated by the glucose molecules present in the aqueous humor, thus carries the blood glucose concentration information. A proof-of-concept experimental bench-top setup is presented, applying a multi-wavelength true phase measurement approach and a rotating phase retarder as an analyzer to measure the very small rotation angles and the complete polarization state of the measurement light.

Paper Details

Date Published: 14 October 2003
PDF: 12 pages
Proc. SPIE 5143, Novel Optical Instrumentation for Biomedical Applications, (14 October 2003); doi: 10.1117/12.502040
Show Author Affiliations
Stefan Boeckle, Univ. degli Studi di Brescia (Italy)
Luigi L. Rovati, Univ. degli Studi di Modena e Reggio Emilia (Italy)
Rafat R. Ansari, NASA Glenn Research Ctr. (United States)