The calibration system used for measurements of the frequency characteristics of 633 nm He-Ne lasers light is presented in the paper. The system consists of four blocks: the iodine stabilized He-Ne at 633 nm laser frequency standard, optoelectronic unit, microprocessor system and PC compatible computer. The presented system was applied for measurements of the frequency characteristics of the frequency stabilized He-Ne lasers used in the interferometric systems. The analysis of the obtained results is discussed in paper.

Paper Details

Date Published: 1 April 2003
PDF: 4 pages
Proc. SPIE 5064, Lightmetry 2002: Metrology and Testing Techniques Using Light, (1 April 2003); doi: 10.1117/12.501518