Share Email Print
cover

Proceedings Paper

OCT system with the fiber optic interferometer and Cr:forsterite laser
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The possibility of using short-pulse laser with interferometric fiber-optic for diagnostic of tissues is presented. The Cr:Forsterite laser operated at 1300 nm is proposed and the first characteristics of such laser are presented. The detection of backscattering light is obtained by using the Michelson type fiber-optic interferometer. The main aspect of signal processing based on a phase calculation as well as theoretical system description in the Jones’ matrix approach is also presented.

Paper Details

Date Published: 2 October 2003
PDF: 8 pages
Proc. SPIE 5140, Optical Coherence Tomography and Coherence Techniques, (2 October 2003); doi: 10.1117/12.501077
Show Author Affiliations
Andrzej S Zajac, Military Univ. of Technology (Poland)
Leszek R Jaroszewicz, Military Univ. of Technology (Poland)
Marek Skorczakowski, Military Univ. of Technology (Poland)
Pawel Marc, Military Univ. of Technology (Poland)


Published in SPIE Proceedings Vol. 5140:
Optical Coherence Tomography and Coherence Techniques
Wolfgang Drexler, Editor(s)

© SPIE. Terms of Use