Characteristics of a high-pressure gas proportional counter filled with xenon
Author(s): Hirohisa Sakurai; Brian D. Ramsey
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

The characteristics of a conventional cylindrical geometry proportional counter filled with high pressure xenon gas up to 10 atm. were fundamentally investigated for use as a detector in hard X-ray astronomy. With a 2 percent methane gas mixture the energy resolutions at 10 atm. were 9.8 percent and 7.3 percent for 22 keV and 60 keV X-rays, respectively. From calculations of the Townsend ionization coefficient, it is shown that proportional counters at high pressure operate at weaker reduced electric field than low pressure counters. The characteristics of a parallel grid proportional counter at low pressure showed similar pressure dependence. It is suggested that this is the fundamental reason for the degradation of resolution observed with increasing pressure.

Paper Details

Date Published: 1 October 1991
PDF: 8 pages
Proc. SPIE 1549, EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy II, (1 October 1991); doi: 10.1117/12.48323