In biomedical imaging high speed axial scanning is very important for obtaining the cross-sectional images of tissue microstructures. For this purpose our optical coherence tomography (OCT) system incorporated a scanning optical delay line (RSOD) that uses Fourier-transform pulse shaping technique. In this technique axial scanning in the time domain is implemented by a linear phase ramp in the frequency domain. By combining the axial scanning (Z-axis) in the reference arm and lateral scanning (X-axis) in the sample arm using a lateral scanning galvanometer, the system could image samples with a size of 400×800 pixels (Z-X image) at the speed of 6 frames per second. Some imaging results are presented in this paper.

Paper Details

Date Published: 12 September 2002
PDF: 7 pages
Proc. SPIE 4916, Optics in Health Care and Biomedical Optics: Diagnostics and Treatment, (12 September 2002); doi: 10.1117/12.482972
Show Author Affiliations
Guiju Song, Oregon Health and Science Univ. (United States)
Steven L. Jacques, Oregon Health and Science Univ. (United States)