Contrast-to-noise and exposure measurements of an aSi:H/Csl(Tl) flat-panel based digital radiography system using a QC chest phantom
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The amorphous silicon (a-Si:H) flat panel imaging systems have recently become commercially available and gained acceptance as promising candidates for implementing digital radiography. In this work, SNRs, CNRs and figure-of-merit, defined as CNR2/SEEs, were measured for a commercial a-Si:H/CsI(Tl) flat-panel digital chest radiography system as a function of the kVp for three different regions in the images of a chest phantom: lung, heart and subphrenic. Using these measurements, issues of image quality and optimal selection of the kVp are discussed.

Paper Details

Date Published: 5 June 2003
PDF: 7 pages
Proc. SPIE 5030, Medical Imaging 2003: Physics of Medical Imaging, (5 June 2003); doi: 10.1117/12.480473