Enhancements which have been made in the Radiometric Calibration Facility (RCF) of the CERES payload are described. These include narrow field blackbody and wide field of view blackbody sources, an active cavity radiometer, and a shortwave reference source. They permit the RCF to calibrate the CERES instruments to better than +/- 0.5 percent absolute radiometric accuracy in the 3.5 to above 50 micron wavelength region and to +/- 1.0 percent in the 0.3-3.5 micron region.

Paper Details

Date Published: 1 August 1991
PDF: 12 pages
Proc. SPIE 1493, Calibration of Passive Remote Observing Optical and Microwave Instrumentation, (1 August 1991); doi: 10.1117/12.46703
Show Author Affiliations
Mark A. Folkman, TRW Space and Technology Group (United States)
Peter J. Jarecke, TRW Space and Technology Group (United States)
Lane A. Darnton, TRW Space and Technology Group (United States)