This article presents results of a theoretical and simulating analysis of the accuracy of the method of the direct measurement of resistance using a charge balancing circuit and the 'autozero' method. Achieved results showed that those construction makes it possible to reduce a majority of factors that could be sources of measurement errors, however an extent of the reduction first of all depends on parameters of the used semiconductor switches. The influence of finite values of on-state resistances and switching times on a linearity and monotonicity of the conversion characteristic were analyzed in the paper. Results of simulations have been based on parameters of contemporary MOSFET transistors that can be used to switch currents in the circuit.

Paper Details

Date Published: 8 August 2001
PDF: 5 pages
Proc. SPIE 4516, Optoelectronic and Electronic Sensors IV, (8 August 2001); doi: 10.1117/12.435918
Show Author Affiliations
J. Gronczynski, Wroclaw Univ. of Technology (Poland)
Janusz Mroczka, Wroclaw Univ. of Technology (Poland)