Share Email Print
cover

Proceedings Paper

Megavoltage radiation dose response of digital computed radiography
Author(s): Zhiheng Wang; Qingrong J. Wu; Claudio H. Sibata
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The response of a digital computed radiography system to the megavoltage therapeutic radiation beams was invested. A narrow slit of radiation beam was used to test the line spread function of the system. The effects of various facts such as cassette, beam energy, radiation dose, scanning orientation and timing on the line spread function were investigated. The calibration curves were established to calibrate the image intensity to the megavoltage radiation dose. The calibration curves were applied to measure the beam profiles of the radiation fields with various wedges.

Paper Details

Date Published: 28 June 2001
PDF: 8 pages
Proc. SPIE 4320, Medical Imaging 2001: Physics of Medical Imaging, (28 June 2001); doi: 10.1117/12.430958
Show Author Affiliations
Zhiheng Wang, Case Western Reserve Univ. School of Medicine (United States)
Qingrong J. Wu, Case Western Reserve Univ. School of Medicine (United States)
Claudio H. Sibata, Case Western Reserve Univ. School of Medicine (United States)


Published in SPIE Proceedings Vol. 4320:
Medical Imaging 2001: Physics of Medical Imaging
Larry E. Antonuk; Martin Joel Yaffe, Editor(s)

© SPIE. Terms of Use