The performance of a digital radiography system that included a prototype flat panel detector (StingRay) was compared with a 400 speed screen-film system. The flat panel detector consisted of a 500 micrometers thick CsI scintillator with an image matrix size of 3k2. The limiting spatial resolution of screen-film (approximately 4 line pairs/mm) was superior to that of the flat panel detector (approximately 2.5 line pairs/mm). The digital detector had an excellent linearity response (r2 equals 0.997), a dynamic range of 20,000:1, and saturated at a radiation exposure of 60 mR.