Paper Abstract

A theory of the imaging behavior of structured phosphor layers is developed based on linear analysis and Carlo model modeling and tested by comparison with experiments. The experiments include: the evaluation of Swank factor; modulation transfer function (MTF) measured as a function of spatial frequency f on CsI phosphor layers of different thicknesses and optical properties (e.g. presence or absence of reflective backing layer). These measurements on structured screens were used to establish phosphor parameters (e.g. absorption and scattering lengths). In addition, MTF(f), Wiener noise power spectra [NPS(f)] and detective quantum efficiency [DQE(f)] measurements were obtained from the literature for x-ray image intensifiers and indirect conversion flat panel detectors constructed using CsI layers. These results were compared with the theory using the previously established phosphor parameters. The effects of K-fluorescence and depth dependence of the MTF on DQE (i.e. Lubberts' effect) were investigated.

Paper Details

Date Published: 28 June 2001
PDF: 11 pages
Proc. SPIE 4320, Medical Imaging 2001: Physics of Medical Imaging, (28 June 2001); doi: 10.1117/12.430931
Show Author Affiliations
John A. Rowlands, Sunnybrook and Women's College Health Sciences Ctr./Univ. of Toronto (Canada)
Winston Guangda Ji, Sunnybrook and Women's College Health Sciences Ctr./Univ. of Toronto (Canada)
Wei Zhao, SUNY/Stony Brook (United States)


Published in SPIE Proceedings Vol. 4320:
Medical Imaging 2001: Physics of Medical Imaging
Larry E. Antonuk; Martin Joel Yaffe, Editor(s)

© SPIE. Terms of Use
Back to Top