Share Email Print
cover

Proceedings Paper

Development of an automated method for evaluation of sharpness of digital radiographs using edge method
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The presampling modulation transfer function (MTF) can be determined by the edge spread function in which the sampling interval is narrower than the pixel-to-pixel interval from slight angled edge image. It is important that the precision of the presampling MTF depend on the precision of the edge angle. In this study, we have developed the automated method, which includes a precise edge angle determination process for the measurement of the presampling MTF.

Paper Details

Date Published: 28 June 2001
PDF: 8 pages
Proc. SPIE 4320, Medical Imaging 2001: Physics of Medical Imaging, (28 June 2001); doi: 10.1117/12.430929
Show Author Affiliations
Ikuo Kawashita, Hiroshima International Univ. (Japan)
Kouji Maeda, Hiroshima International Univ. (Japan)
Hidetaka Arimura, Hiroshima International Univ. (Japan)
Kaoru Morikawa, Hiroshima International Univ. (Japan)
Takayuki Ishida, Hiroshima International Univ. (Japan)


Published in SPIE Proceedings Vol. 4320:
Medical Imaging 2001: Physics of Medical Imaging
Larry E. Antonuk; Martin Joel Yaffe, Editor(s)

© SPIE. Terms of Use