DQE(f) of an amorphous-silicon flat-panel x-ray detector: detector parameter influences and measurement methodology
Author(s): Paul R. Granfors; Richard Aufrichtig
Format Member Price Non-Member Price
PDF $17.00 $21.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

We discuss how the frequency dependent detective quantum efficiency [DQE(f)] in a well-designed amorphous silicon flat panel detector is affected by several phenomena that reduce the DQE in other types of medical imaging detectors. The detector examined employs a CsI(Tl) scintillator and is designed for general diagnostic x-ray imaging applications. We consider DQE degradation due to incomplete x-ray absorption, secondary quantum noise, Swank factor, Lubberts effect, spatial variation in gain, noise aliasing, and additive system noise. The influences of detector design parameters on the frequency- and exposure-dependent DQE are also examined. We find that the DQE does not depend directly on MTF and that DQE is independent of exposure within the detector's operating range, except at the lowest exposures. Likewise the signal per absorbed x-ray, which contains the fill factor as one of several multiplicative components, does not affect DQE except at the lowest exposures. A methodology for determining DQE(f) from measurements of MTF(f), noise power spectrum (NPS), average signal, and x-ray exposure is presented. We find that it is important to incorporate several corrections in the NPS measurement procedure in order to obtain accurate results. These include corrections for lag, non-linearity, response variation from pixel to pixel, and use of a finite number of flat-field images. MTF, NPS, and DQE results are presented for a 41 X 41-cm2 flat panel detector designed for radiographic applications.

Paper Details

Date Published: 25 April 2000
PDF: 12 pages
Proc. SPIE 3977, Medical Imaging 2000: Physics of Medical Imaging, (25 April 2000); doi: 10.1117/12.384481
Show Author Affiliations
Paul R. Granfors, GE Medical Systems (United States)
Richard Aufrichtig, GE Medical Systems (United States)


Published in SPIE Proceedings Vol. 3977:
Medical Imaging 2000: Physics of Medical Imaging
James T. Dobbins III; John M. Boone, Editor(s)