X-ray and gamma-ray spectroscopy with segmented CdZnTe detectors
Author(s): A. Shor; Yossi Eisen; Israel D. Mardor
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Two methods are presented for obtaining excellent spectroscopic performance and good photo-peak efficiencies for CdZnTe coarsely segmented pad and strip detectors. The `correlation method' utilizes the correlation between the signals in the planar and segmented electrodes. This correlation provides an indication of the point of interaction. Application of a correction which is based on this correlation generates spectra with a resolution of 1.4% FWHM at 356 keV for pad detectors. Furthermore, for strip detectors this method yields improved spectra and an increase of more than 35% in photo-peak efficiencies at 140.5 keV. The `compensation method' involves tuning the degree of electron trapping to compensate for the almost completely trapped holes. Contrary to planar detectors, for segmented detectors the compensation condition yields induced charge signals which are 80% of the maximum signal. Using this method we obtained a spectrum with energy resolution of 3.5% FWHM at 122 keV and a peak-to-valley ratio of approximately 100:1.

Paper Details

Date Published: 1 October 1999
PDF: 12 pages
Proc. SPIE 3769, Penetrating Radiation Systems and Applications, (1 October 1999); doi: 10.1117/12.363692
Show Author Affiliations
A. Shor, Soreq Nuclear Research Ctr. (Israel)
Yossi Eisen, Soreq Nuclear Research Ctr. (Israel)
Israel D. Mardor, Soreq Nuclear Research Ctr. (Israel)


Published in SPIE Proceedings Vol. 3769:
Penetrating Radiation Systems and Applications
F. Patrick Doty, Editor(s)

© SPIE. Terms of Use
Back to Top