Share Email Print
cover

Proceedings Paper

Design and stability measurement of 543-nm He-Ne-I2 laser system
Author(s): Josef Lazar
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

In this contribution, the design of an iodine stabilized laser system is presented. The arrangement is with an external iodine cell, and a detection system employs a derivative spectroscopy technique based on the third harmonic phase sensitive demodulation. To prove the frequency stability, the comparison with a 633 nm He-Ne-I2 stabilized laser was arranged. To record a beat signal frequency, an additional 633 nm He-Ne laser which tracks frequency fluctuations of the 543 nm He-Ne-I2 system was used.

Paper Details

Date Published: 14 July 1999
PDF: 4 pages
Proc. SPIE 3820, 11th Slovak-Czech-Polish Optical Conference on Wave and Quantum Aspects of Contemporary Optics, (14 July 1999); doi: 10.1117/12.353074
Show Author Affiliations
Josef Lazar, Institute of Scientific Instruments (Czech Republic)


Published in SPIE Proceedings Vol. 3820:
11th Slovak-Czech-Polish Optical Conference on Wave and Quantum Aspects of Contemporary Optics
Miroslav Hrabovsky; Anton Strba; Waclaw Urbanczyk, Editor(s)

© SPIE. Terms of Use