Share Email Print
cover

Proceedings Paper

System of measurement of the transmission spectrum of in-vitro corneas for eye banks
Author(s): Josemilson de Meneses Bispo; Liliane Ventura; Carlos Roberto de Carvalho; Sidney Julio Faria de Sousa; Caio Chiaradia
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The donated cornea transparency is one of the features to be analyzed previously to its indication for the transplant. Hence we have developed a system to evaluate the transmission spectrum of 'In Vitro' corneas in its preservative medium, in the range of 400 - 700 nm, to be implemented in an Eye Bank. The cornea in its preservative medium is illuminated by collimated white light beams which strike a diffraction grid, passes through an optical system and a selecting filter (400 - 800 nm) and then is detected by a linear CCD detector (2048 photodiodes). The data is displayed in a PC monitor via a commercial interface device. A dedicated software has been developed to plot the transmission spectrum profile and provide an objective and standard information about the transparency of the donated cornea.

Paper Details

Date Published: 13 May 1998
PDF: 4 pages
Proc. SPIE 3254, Laser-Tissue Interaction IX, (13 May 1998); doi: 10.1117/12.308185
Show Author Affiliations
Josemilson de Meneses Bispo, Univ. de Sao Paulo (Brazil)
Liliane Ventura, Univ. de Sao Paulo (Brazil)
Carlos Roberto de Carvalho, Univ. de Sao Paulo (Brazil)
Sidney Julio Faria de Sousa, Univ. de Sao Paulo (Brazil)
Caio Chiaradia, Univ. de Sao Paulo (Brazil)


Published in SPIE Proceedings Vol. 3254:
Laser-Tissue Interaction IX
Steven L. Jacques; Jeff Lotz, Editor(s)

© SPIE. Terms of Use