Share Email Print
cover

Proceedings Paper

Sol gel-based fiber optic pH sensor
Author(s): Sheila A. Grant; Robert S. Glass
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

We report here on the development of a sol-gel based fiber optic sensor to monitor local blood pH. These sensors are being developed for invasive medical applications; that is, they will be coaxially threaded through a catheter beyond an occlusion in the vascular system. The fiber optic pH sensor design was based on the immobilization of a pH sensitive dye, seminaphthorhodamine-1 carboxylate, onto the tip or surface of an optical fiber using the sol-gel method. The fiber optic pH sensor was tested in phosphate buffer saline and human whole blood using a miniature fluorimeter system. Linear responses were obtained in blood in the pH range 6.7 to 8.0, which demonstrates potential for use for in vivo sensing.

Paper Details

Date Published: 16 June 1997
PDF: 7 pages
Proc. SPIE 2976, Biomedical Sensing, Imaging, and Tracking Technologies II, (16 June 1997); doi: 10.1117/12.275514
Show Author Affiliations
Sheila A. Grant, Lawrence Livermore National Lab. (United States)
Robert S. Glass, Lawrence Livermore National Lab. (United States)


Published in SPIE Proceedings Vol. 2976:
Biomedical Sensing, Imaging, and Tracking Technologies II
Tuan Vo-Dinh; Robert A. Lieberman; Gerald G. Vurek; Abraham Katzir, Editor(s)