Comparison of computer analysis of mammography phantom images (CAMPI) with perceived image quality of phantom targets in the ACR phantom
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

Computer analysis of mammography phantom images (CAMPI) is a method for making quantitative measurements of image quality on phantom images. The purpose of this work was to determine the correlation of the CAMPI measures with the perceived image quality of the microcalcification target objects in a phantom. A large existing phantom image database was subjected to CAMPI analysis. Extracted microcalcification regions from some of these images were compared in pairs by three observers. A global maximization technique was used to determine which linear combinations of the CAMPI measures most closely predicted the paired comparison observations. An analysis was also conducted to determine the validity of the linear model. It was found that the signal-to-noise- ratio measure (SNR) and the correlation measure (COR) most strongly correlated with the observed paired comparison results. A linear combination of the measures gave a slightly, but significantly, better correlation with the observations.

Paper Details

Date Published: 16 April 1997
PDF: 8 pages
Proc. SPIE 3036, Medical Imaging 1997: Image Perception, (16 April 1997); doi: 10.1117/12.271288