Optimized conversion from CT numbers to proton relative stopping power based on proton radiography and scatter corrected cone-beam CT images
Author(s): Nils Krah; Simon Rit
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

We propose a method to generate accurate proton relative stopping power (RSP) maps from patient cone-beam CT (CBCT) images. The scatter polluted low frequency component in the CBCT projections is replaced by an analytically calculated estimate of the scatter-free component. This is obtained by forward projecting the segmented CBCT image overridden with reference materials (air, soft tissue, bone). The projection model accounts for polychromaticity and uses an estimate of the combined source-detector spectral function. High and low frequency image components are automatically matched. The accurate conversion curve from CT numbers to RSP is obtained by comparison of a proton radiography and a proton digitally reconstructed radiography. CBCT images are acquired on a clinical scanner and proton images are simulated by Monte Carlo. Results show clearly reduced cupping effect and overall better RSP accuracy when CBCT images are scatter corrected.

Paper Details

Date Published: 28 May 2019
PDF: 4 pages
Proc. SPIE 11072, 15th International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, 110720Y (28 May 2019); doi: 10.1117/12.2534898
Show Author Affiliations
Nils Krah, Univ. Lyon, CREATIS, CNRS (France)
Simon Rit, Univ. Lyon, CREATIS, CNRS (France)


Published in SPIE Proceedings Vol. 11072:
15th International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine
Samuel Matej; Scott D. Metzler, Editor(s)