It has developed and performed a novel fiberoptic immunosensor instrument with operating wavelength 400 - 760 nm and repeatability cv equals 0.27%. The instrument has many excellent features such as simplified operation, faster testing time, higher sensitivity and economic cost. It has completely eliminated recovery period which traditional immunosensor owned due to use separative sensor structure. It can widely apply to test for bacteria, virus, hormone, parasite and cancer protein in clinical examination. The instrument has operated in laboratory and relevant medicine units and successfully tested monoclonal rat-anti-human of 413 cases in clinic and prepared with existing ELISA method, the coincidence probability reached 94 to 100%.