Performance comparison of water phantom based flat field correction methods for photon-counting spectral CT images: experimental results
Author(s): Donghyeok Kim; Jongduk Baek
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Photon-counting detector (PCD) is one of the leading candidates for the next generation X-ray detector. The PCD enables effective multiple material decomposition and low dose CT imaging, but several correction techniques to compensate non-ideal effects of the PCD should be preceded to maximize the imaging performance. In this study, we present experimental results of water phantom based flat field correction methods to reduce energy distortion for photon-counting spectral CT images. We compared the performance of the flat field correction using single and multiple water phantoms, and examined the robustness of the correction method for various imaging tasks. The flat field correction using single water phantom was conducted by normalizing measured projection data of an object with those of the centered large water phantom. Then, ideal water phantom image was added to the reconstructed image of the water normalized projection data, which is equivalent to the conducting first order fitting of detector pixel gain. The dynamic range of detector pixel gain was increased by using multiple water phantoms. Each detector pixel gain was modeled using 2nd order polynomial, and the coefficients were estimated by comparing measured projection data of water phantoms with those of ideal water phantoms. The estimated gain of each detector pixel was tested for various imaging tasks. The results show that the single water phantom based method reduces the ring artifacts for a centered object effectively, but residual ring artifacts are introduced for the off-centered multiple water phantoms. In contrast, the multiple water phantoms based method reduces the ring artifacts effectively regardless of the size and location of the object. In addition, the multiple water phantoms based method provides improved SNR by 38.3 to 41.9% and CNR by 52.8 to 56.5%, compared to the single water phantom based method.

Paper Details

Date Published: 1 March 2019
PDF: 7 pages
Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 109484K (1 March 2019); doi: 10.1117/12.2512699
Show Author Affiliations
Donghyeok Kim, Yonsei Univ. (Korea, Republic of)
Samsung Electronics Co., Ltd. (Korea, Republic of)
Jongduk Baek, Yonsei Univ. (Korea, Republic of)


Published in SPIE Proceedings Vol. 10948:
Medical Imaging 2019: Physics of Medical Imaging
Taly Gilat Schmidt; Guang-Hong Chen; Hilde Bosmans, Editor(s)

© SPIE. Terms of Use
Back to Top