Share Email Print
cover

Proceedings Paper • new

OCT imaging along retinal vessels using arbitrary scanning patterns (Conference Presentation)
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Investigations on retinal vasculature and blood flow are of interest for understanding and diagnostics of numerous ocular diseases. Conventional OCT systems use various scan patterns like linear B-scans, circular scans around the optic nerve head, or raster scans for 3D data acquisition. However, for some studies it is preferable to have customized scan patterns that can, e.g., follow the trace of an arbitrary linear structure in the retina, such as a vessel. In this work, we present an OCT instrument with an integrated retinal tracker that allows repeated scans along an arbitrary trace, whereby ocular motions are corrected by the retinal tracker. The setup comprises an OCT system and a line scanning laser ophthalmoscope (LSLO). The OCT subsystem operates at a center wavelength of 860 nm, with a bandwidth of 60 nm and an A-scan rate of 70kHz. The LSLO system operates at 790 nm and at a frame rate of 60 Hz. This system was used for reflectivity and Doppler imaging along retinal vessels. In a first step, the vessel is manually marked on the LSLO image. Then, repeated scans along the vessel trace are performed (2048 A-scans per scan along trace, up to 500 scans along the trace). The intensity images show a clear delineation of the vessel walls, the phase difference (Doppler) tomograms allow for a time-resolved analysis of blood flow along the vessel over the cardiac cycle.

Paper Details

Date Published: 4 March 2019
PDF
Proc. SPIE 10867, Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XXIII, 108671B (4 March 2019); doi: 10.1117/12.2511117
Show Author Affiliations
Sylvia Desissaire, Medizinische Univ. Wien (Austria)
Andreas Wartak, Medizinische Univ. Wien (Austria)
Massachusetts General Hospital, Harvard Medical School (United States)
Florian Beer, Medizinische Univ. Wien (Austria)
Matthias Salas, Medizinische Univ. Wien (Austria)
Mitsuro Sugita, Medizinische Univ. Wien (Austria)
Bernhard Baumann, Medizinische Univ. Wien (Austria)
Michael Pircher, Medizinische Univ. Wien (Austria)
Christoph K. Hitzenberger, Medizinische Univ. Wien (Austria)


Published in SPIE Proceedings Vol. 10867:
Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XXIII
James G. Fujimoto; Joseph A. Izatt, Editor(s)

© SPIE. Terms of Use
Back to Top