ROSAT WFC imaging detectors
Author(s): Martin Adrian Barstow; Anne E. Sansom
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Results of the calibration program performed on flight and flight-spare detectors for the Rosat Wide Field Camera (WFC) are presented. The result of an accelerated life test on a development model detector assembled to flight standard are summarized. Imaging tests demonstrate that the lookup table technique for removing distortion works efficiency with low differential nonlinearity. No undesirable 'chicken wire' effects are seen in the images, and the detector resolution matches the on-axis performance of the telescope and is constant across the field of view. Peaks in efficiency occur at 10.2, 20, and 100 eV and mimima at 13 and 45 eV. The secondary 13 eV minimum is correlated with the onset of two-electron photoemission. The mean change in gain as a function of photon energy in the EUV band is much less rapid than in the soft X-ray band.

Paper Details

Date Published: 1 November 1990
PDF: 11 pages
Proc. SPIE 1344, EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy, (1 November 1990); doi: 10.1117/12.23253
Show Author Affiliations
Martin Adrian Barstow, Univ. of Leicester (United Kingdom)
Anne E. Sansom, Univ. of Leicester (United Kingdom)


Published in SPIE Proceedings Vol. 1344:
EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy