Impact of photon counting detector technology on kV selection and diagnostic workflow in CT
Author(s): Wei Zhou; Dilbar Abdurakhimova; Michael Bruesewitz; Ahmed Halaweish; Cynthia H. McCollough; Shuai Leng
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The purpose of this study is to determine the optimal iodine contrast-to-noise ratio (CNR) achievable for different patient sizes using virtual-monoenergetic-images (VMIs) and a universal acquisition protocol on photon-counting-detector CT (PCD-CT), and to compare results to those from single-energy (SE) and dualsource- dual-energy (DSDE) CT. Vials containing 3 concentrations of iodine were placed in torso-shaped water phantoms of 5 sizes and scanned on a 2nd generation DSDE scanner with both SE and DE modes. Tube current was automatically adjusted based on phantom size with CTDIvol ranging from 5.1 to 22.3 mGy. PCD-CT scans were performed at 140 kV, 25 and 75 keV thresholds, with CTDIvol matched to the SE scans. DE VMIs were created and CNR was calculated for SE images and DE VMIs. The optimal kV (SE) or keV (DE VMI) was chosen at the point of highest CNR with no noticeable artifacts. For 10 mgI/cc vials in the 35 cm phantom, the optimal CNR of VMIs on PCD (22.6@50keV) was comparable to that of the best DSDE protocol (23.9@50keV) and was higher than that of the best SE protocol (19.7@80kV). In general, the difference of optimal CNR between PCD and SE increased with phantom size, with PCD 50 keV VMIs having an equivalent CNR (0.6% difference) with that of SE at the 25 cm phantom and 57% higher CNR at the 45 cm phantom. PCD-CT demonstrated comparable iodine CNR of VMIs to that of DSDE across patient sizes. Whereas SE and DSDE CT exams require use of patient-size-specific acquisitions settings, our findings point to the ability of PCD-CT to simplify protocol selection, using a single VMI keV setting (50 keV), acquisition kV (140 kV), and energy thresholds (25 and 75 keV) for all patient sizes, while achieving optimal or near optimal iodine CNR values.

Paper Details

Date Published: 9 March 2018
PDF: 7 pages
Proc. SPIE 10573, Medical Imaging 2018: Physics of Medical Imaging, 105731C (9 March 2018); doi: 10.1117/12.2294952
Show Author Affiliations
Wei Zhou, Mayo Clinic (United States)
Dilbar Abdurakhimova, Mayo Clinic (United States)
Michael Bruesewitz, Mayo Clinic (United States)
Ahmed Halaweish, Siemens Healthcare (United States)
Cynthia H. McCollough, Mayo Clinic (United States)
Shuai Leng, Mayo Clinic (United States)


Published in SPIE Proceedings Vol. 10573:
Medical Imaging 2018: Physics of Medical Imaging
Joseph Y. Lo; Taly Gilat Schmidt; Guang-Hong Chen, Editor(s)

© SPIE. Terms of Use
Back to Top