Share Email Print
cover

Proceedings Paper

Optimized modeling of powder-based x-ray scintillator screen for improved total efficiency by double-layer scintillator design (Conference Presentation)

Paper Abstract

Powder-based flat panel X-ray scintillator screen has been widely used in many medical fields for diagnosis and measurement. For the proper application in medical field, high dose usage and high image quality are required. In this study, double-layer scintillator design is suggested for improving total efficiency with using theoretical x-ray absorption model and Mie-scattering-based Monte-Carlo (LightTools) analysis. Optimized scintillator layer design would be proposed with respect to the medical fields’ sufficient resolution-functionality.

Paper Details

Date Published: 25 September 2017
PDF
Proc. SPIE 10393, Radiation Detectors in Medicine, Industry, and National Security XVIII, 103930N (25 September 2017); doi: 10.1117/12.2273566
Show Author Affiliations
Jookwon Song, Yonsei Univ. (Korea, Republic of)
Hyeon Bo Shim, Yonsei Univ. (Korea, Republic of)
Jae W. Hahn, Yonsei Univ. (Korea, Republic of)


Published in SPIE Proceedings Vol. 10393:
Radiation Detectors in Medicine, Industry, and National Security XVIII
Gary P. Grim; Lars R. Furenlid; H. Bradford Barber, Editor(s)

© SPIE. Terms of Use