Study of the optical crosstalk in a heterodyne displacement gauge with cancelable circuit
Author(s): Alberto Donazzan; Giampiero Naletto; Maria G. Pelizzo
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

One main focus of high precision heterodyne displacement interferometers are the means of splitting and merging for the reference (R) and measurement (M) beams when a cancelable circuit is implemented. Optical mixing of R and M gives birth to a systematc error called cyclic error, which appears as a periodic offset between the detected displacement and the actual one.

A simple derivation of the cyclic error due to optical mixing is proposed for the cancelable circuit design. R and M beatings are collected by two photodiodes and conveniently converted by transimpedance amplifiers, such that the output signals are turned into ac-coupled voltages. The detected phase can be calculated as a function of the real phase (a change in optical path difference) in the case of zero-crossing detection. What turns out is a cyclic non-linearity which depends on the actual phase and on the amount of optical power leakage from the R channel into the M channel and vice versa. We then applied this result to the prototype of displacement gauge we are developing, which implements the cancelable circuit design with wavefront division. The splitting between R and M is done with a double coated mirror with a central hole, tilted by 45° with respect to the surface normal. The interferometer features two removable diffraction masks, respectively located before the merging point (a circular obscuration) and before the recombination point (a ring obscuration). In order to predict the extent of optical mixing between R and M, the whole layout was simulated by means of the Zemax ® Physical Optics Propagation (POP) tool. After the model of our setup was built and qualitatively verified, we proceeded by calculating the amount of optical leakages in various configurations: with and without the diffraction masks as well as for different sizes of both the holey mirror and the diffraction masks. The corrisponding maximum displacement error was then calculated for every configuration thanks to the previously derived formula. The insertion and optimization of the diffraction masks greatly improved the expected optical isolation inside the system.

Data acquisition from our displacement gauge has just started. We plan to experimentally verify such results as soon as our prototype gauge will reach the desired sub-nanometer sensitivity.

Paper Details

Date Published: 26 June 2017
PDF: 6 pages
Proc. SPIE 10330, Modeling Aspects in Optical Metrology VI, 103300F (26 June 2017); doi: 10.1117/12.2270193
Show Author Affiliations
Alberto Donazzan, Univ. degli Studi di Padova (Italy)
Istituto Nazionale di Fisica Nucleare, Sezione di Padova (Italy)
Giampiero Naletto, Univ. degli Studi di Padova (Italy)
Istituto Nazionale di Fisica Nucleare, Sezione di Padova (Italy)
CNR-IFN, Sezione di Padova (Italy)
Maria G. Pelizzo, Istituto Nazionale di Fisica Nucleare, Sezione di Padova (Italy)
CNR-IFN, Sezione di Padova (Italy)


Published in SPIE Proceedings Vol. 10330:
Modeling Aspects in Optical Metrology VI
Bernd Bodermann; Karsten Frenner; Richard M. Silver, Editor(s)

© SPIE. Terms of Use
Back to Top