This paper presents a contactless method for gauge blocks calibration combining laser interferometry and low-coherence interferometry. In the presented system, the contactless measurement of the absolute gauge block length is done as a single-step operation without any change in optical setup during measurement, giving complete information about the gauge block length. The paper also presents a set of optimization steps which have been done in order to transform the original experimental setup into the automatic system which meets legal length metrology requirements. To prove the measurement traceability, we conducted a set of gauge block length measurement comparing data from the optimized system and the established reference systems TESA NPL A.G.I. 300 and TESA–UPC operated in Czech Metrology Institute laboratory.

Paper Details

Date Published: 23 December 2016
PDF: 10 pages
Proc. SPIE 10142, 20th Slovak-Czech-Polish Optical Conference on Wave and Quantum Aspects of Contemporary Optics, 101420O (23 December 2016); doi: 10.1117/12.2263367
Show Author Affiliations
Zdeněk Buchta, Institute of Scientific Instruments of the ASCR, v.v.i. (Czech Republic)
Martin Šarbort, Institute of Scientific Instruments of the ASCR, v.v.i. (Czech Republic)
Martin Čížek, Institute of Scientific Instruments of the ASCR, v.v.i. (Czech Republic)
Václav Hucl, Institute of Scientific Instruments of the ASCR, v.v.i. (Czech Republic)
Šimon Řeřucha, Institute of Scientific Instruments of the ASCR, v.v.i. (Czech Republic)
Tomáš Pikálek, Institute of Scientific Instruments of the ASCR, v.v.i. (Czech Republic)
Štěpánka Dvořáčková, Technical Univ. of Liberec (Czech Republic)
František Dvořáček, Czech Metrology Institute (Czech Republic)
Jan Kůr, MESING Brno a.s. (Czech Republic)
Pavel Konečný, MESING Brno a.s. (Czech Republic)
Josef Lazar, Institute of Scientific Instruments of the ASCR, v.v.i. (Czech Republic)
Ondřej Číp, Institute of Scientific Instruments of the ASCR, v.v.i. (Czech Republic)