Share Email Print
cover

Proceedings Paper

Development of a novel method based on a photon counting technique with the aim of precise material identification in clinical x-ray diagnosis
Author(s): Natsumi Kimoto; Hiroaki Hayashi; Takashi Asahara; Yuki Kanazawa; Tsutomu Yamakawa; Shuichiro Yamamoto; Masashi Yamasaki; Masahiro Okada
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

A photon counting system has the ability of energy discrimination, therefore obtaining new information using X-rays for material identification is an expected goal to achieve precise diagnosis. The aim of our study is to propose a novel method for material identification based on a photon counting technique. First, X-ray spectra at 40-60 kV were constructed using a published database. Second, X-ray spectra penetrating different materials having atomic numbers from 5-13 were calculated. These spectra were divided into two energy regions, then linear attenuation factors concerning these regions were obtained. In addition, in order to accomplish highly accurate material identification, correction of beam hardening effects based on soft-tissue was applied to each linear attenuation factor. Then, using the linear attenuation factors, a normalized linear attenuation coefficient was derived. Finally, an effective atomic number was determined using the theoretical relationship between the normalized linear attenuation coefficient and atomic number. In order to demonstrate our method, four different phantoms (acrylic, soft-tissue equivalent, bone equivalent, and aluminum) were measured using a single-probe-type CdTe detector under the assumption that the response of the single-probe-type CdTe detector is equal to the response of one pixel of a multi-pixel-type photon counting detector. Each of these phantoms can be completely separated using our method. Furthermore, we evaluated an adoptive limit of beam hardening correction. We found that the adoptive limit depends on the mass thickness and atomic number. Our vision is to realize highly accurate identification for material with narrow range in atomic number.

Paper Details

Date Published: 9 March 2017
PDF: 11 pages
Proc. SPIE 10132, Medical Imaging 2017: Physics of Medical Imaging, 1013239 (9 March 2017); doi: 10.1117/12.2253564
Show Author Affiliations
Natsumi Kimoto, Tokushima Univ. (Japan)
Hiroaki Hayashi, Tokushima Univ. (Japan)
Takashi Asahara, Tokushima Univ. (Japan)
Yuki Kanazawa, Tokushima Univ. (Japan)
Tsutomu Yamakawa, JOB Corp. (Japan)
Shuichiro Yamamoto, JOB Corp. (Japan)
Masashi Yamasaki, JOB Corp. (Japan)
Masahiro Okada, JOB Corp. (Japan)


Published in SPIE Proceedings Vol. 10132:
Medical Imaging 2017: Physics of Medical Imaging
Thomas G. Flohr; Joseph Y. Lo; Taly Gilat Schmidt, Editor(s)

© SPIE. Terms of Use
Back to Top