Xenon arc lamp spectral radiance modelling for satellite instrument calibration
Author(s): Stephen Rolt; Paul Clark; Jürgen Schmoll; Benjamin J. R. Shaw
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

Precise radiometric measurements play a central role in many areas of astronomical and terrestrial observation. We focus on the use of continuum light sources in the absolute radiometric calibration of detectors in an imaging spectrometer for space applications. The application, in this instance, revolves around the ground based calibration of the Sentinel-4/UVN instrument. This imaging spectrometer instrument is expected to be deployed in 2019 and will make spatially resolved spectroscopic measurements of atmospheric chemistry.

The instrument, which operates across the UV/VIS and NIR spectrum from 305-775 nm, is designed to measure the absolute spectral radiance of the Earth and compare it with the absolute spectral irradiance of the Sun. Of key importance to the fidelity of these absolute measurements is the ground based calibration campaign. Continuum lamp sources that are temporally stable and are spatially well defined are central to this process. Xenon short arc lamps provide highly intense and efficient continuum illumination in a range extending from the ultra-violet to the infra-red and their spectrum is well matched to this specific application. Despite their widespread commercial use, certain aspects of their performance are not well documented in the literature.

One of the important requirements in this calibration application is the delivery of highly uniform, collimated illumination at high radiance. In this process, it cannot be assumed that the xenon arc is a point source; the spatial distribution of the radiance must be characterised accurately. We present here careful measurements that thoroughly characterise the spatial distribution of the spectral radiance of a 1000W xenon lamp. A mathematical model is presented describing the spatial distribution. Temporal stability is another exceptionally important requirement in the calibration process. As such, the paper also describes strategies to re-inforce the temporal stability of the lamp output by means of a closed loop attenuation scheme and current control.

Paper Details

Date Published: 29 July 2016
PDF: 11 pages
Proc. SPIE 9904, Space Telescopes and Instrumentation 2016: Optical, Infrared, and Millimeter Wave, 99044V (29 July 2016); doi: 10.1117/12.2232299
Show Author Affiliations
Stephen Rolt, Durham Univ. (United Kingdom)
Paul Clark, Durham Univ. (United Kingdom)
Jürgen Schmoll, Durham Univ. (United Kingdom)
Benjamin J. R. Shaw, Durham Univ. (United Kingdom)


Published in SPIE Proceedings Vol. 9904:
Space Telescopes and Instrumentation 2016: Optical, Infrared, and Millimeter Wave
Howard A. MacEwen; Giovanni G. Fazio; Makenzie Lystrup; Natalie Batalha; Nicholas Siegler; Edward C. Tong, Editor(s)

© SPIE. Terms of Use
Back to Top