Scatter estimation and removal of anti-scatter grid-line artifacts from anthropomorphic head phantom images taken with a high resolution image detector
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

In radiography, one of the best methods to eliminate image-degrading scatter radiation is the use of anti-scatter grids. However, with high-resolution dynamic imaging detectors, stationary anti-scatter grids can leave grid-line shadows and moiré patterns on the image, depending upon the line density of the grid and the sampling frequency of the x-ray detector. Such artifacts degrade the image quality and may mask small but important details such as small vessels and interventional device features. Appearance of these artifacts becomes increasingly severe as the detector spatial resolution is improved. We have previously demonstrated that, to remove these artifacts by dividing out a reference grid image, one must first subtract the residual scatter that penetrates the grid; however, for objects with anatomic structure, scatter varies throughout the FOV and a spatially differing amount of scatter must be subtracted. In this study, a standard stationary Smit-Rontgen X-ray grid (line density - 70 lines/cm, grid ratio - 13:1) was used with a high-resolution CMOS detector, the Dexela 1207 (pixel size - 75 micron) to image anthropomorphic head phantoms. For a 15 x 15cm FOV, scatter profiles of the anthropomorphic head phantoms were estimated then iteratively modified to minimize the structured noise due to the varying grid-line artifacts across the FOV. Images of the anthropomorphic head phantoms taken with the grid, before and after the corrections, were compared demonstrating almost total elimination of the artifact over the full FOV. Hence, with proper computational tools, antiscatter grid artifacts can be corrected, even during dynamic sequences.

Paper Details

Date Published: 31 March 2016
PDF: 10 pages
Proc. SPIE 9783, Medical Imaging 2016: Physics of Medical Imaging, 978364 (31 March 2016); doi: 10.1117/12.2216833
Show Author Affiliations
R. Rana, Toshiba Stroke and Vascular Research Ctr. (United States)
A. Jain, Toshiba Stroke and Vascular Research Ctr. (United States)
A. Shankar, Toshiba Stroke and Vascular Research Ctr. (United States)
D. R. Bednarek, Toshiba Stroke and Vascular Research Ctr. (United States)
S. Rudin, Toshiba Stroke and Vascular Research Ctr. (United States)


Published in SPIE Proceedings Vol. 9783:
Medical Imaging 2016: Physics of Medical Imaging
Despina Kontos; Thomas G. Flohr, Editor(s)

© SPIE. Terms of Use
Back to Top